Why does the controller
overshoot the set point?
1. Pump is too large
Calculate the flow rate required. The metering pump should
be set for no more than 1.5 times the expected depletion
rate of the chemical. Either adjust the flow rate of the
pump down to this level, or replace the pump with one that
better matches the required flow rate.
2. Pump is siphoning
Install anti-siphon valve on metering pump, or on metering
pump discharge line. Avoid "down hill" tubing runs. A
metering pump injecting into open atmosphere can pump more
than its rated capacity if an injection check valve is not
installed.
3. Poor mixing/large lag time
Improve mixing, or move the location of the sensor and/or
replenishment so that chemical additions reach the sensor
faster. If using a flow-through sensor, make sure that the
flow rate through the sampling system is sufficient, (less
than 1 minute to detect additions). If the lag time
can’t be improved, then in pH/ORP applications, a
proportional controller should be used. For other
applications, it may be possible to use the optional 4-20
mA output for proportional control.
4. Fouled sensor
If the response time of the sensor to a step change in
concentration is slow, then the sensor may need to be
cleaned, calibrated or replaced. Follow the sensor
troubleshooting instructions for that particular sensor.
5. Programmed incorrectly
For on/off controllers, make sure the value of the dead
band menu is not too large. For proportional controllers,
increase the value of the proportional band menu.
6. Relay stuck closed
Use the "Hand-Off-Auto" menu to control the relay. If an
ohmmeter doesn’t show the contact opening and
closing, then replace the power supply board or have the
unit serviced.
