pH/ORP
calibration
As a rule, follow the procedure shown in the pH
meter’s instruction manual. These procedures will
vary depending on whether the meter is a simple type with
manual adjustments, a microprocessor type or a pH
transmitter.
Frequency of
calibration
The frequency of calibration is a function of many factors.
These factors include:
1. The accuracy required by the application.
2. The value of the off-specification product versus the
cost of calibration.
3. The coating or abrasive nature of the application.
4. The stability of the pH electrode and pH meter as a
system.
The frequency of calibration is really determined by
experience. At a new installation, calibration might
initially be checked every few hours or shift with the
calibration changes noted in a log. As a pattern of longer
stability is found, the time between calibration checks can
be increased to once a day or once a week. Although the
frequency of calibration is solely the responsibility of
the user, once a week is the longest recommended interval
between calibrations.
System calibration
concepts
The pH electrode and the pH meter should always be
calibrated as a system. Electronic calibration of a pH
meter with a pH signal simulator checks the meter only and
does not correct for imperfections of the pH electrode.
Even if perfect when new, the performance of pH electrodes
varies with time, usually in an unpredictable way. When
changing electrodes or connecting an electrode to a
different pH meter, re-calibration must be performed.
Single point
calibrations
Single point calibrations involve the use of one pH buffer.
They are the easiest to make but can provide misleading
results. They should only be used for quick checking from
time to time.
Two point
calibrations
As their name implies, 2 point calibrations use 2 pH
buffers: for example, buffers 7.00 and 4.00 or buffers 7.00
and 10.00. Two point calibrations correct for the pH
electrode’s offset and span errors. Since both the
offset and span vary with time, the 2 point method is the
preferred one.
Grab sample
calibrations
The grab sample calibration method is used when it is
difficult or undesirable to remove an electrode from a
system. This method involves obtaining a sample of the
liquid being measured and noting the meter’s reading
at that time. The sample’s reading is obtained by use
of a calibrated lab or portable meter and that reading is
compared to that of the on-line meter. The online meter is
adjusted by the difference between the readings. It is
important to use the difference between the readings
because the system’s reading may have changed in the
intervening time. It is important that the sample being
measured by the lab meter be at the process temperature or
erroneous results may occur.
Calibration
procedures
Stepwise calibration procedures are noted in the pH
meter’s instruction manual. The following suggestions
will help make calibrations as accurate as possible:
1. Before placing the electrode in a new buffer, use an
absorbent paper towel or clean absorbent cloth to blot, not
wipe, off any liquid that clings to the electrode. This
action will minimise carry-over that could contaminate the
buffer.
2. Always use fresh buffers. Safely dispose of the buffers
after they have been used for calibration. Do not return
them to their bottles; this action could contaminate the
buffers.
3. Stir the electrode in the buffer to make certain that
the fresh buffer quickly reaches the electrode’s
measuring surface.
